CVE-2014-5432

Vulnérabilité non évaluée dans SIGMA Spectrum Infusion System (Baxter)

CVSS 0.0 · Non évaluée
EPSS 2.6%
Baxter•SIGMA Spectrum Infusion System•Versions : 6.05 (model 35700BAX) with wireless battery module (WBM) version 16.

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Gravité et risque d'exploitation

CVE-2014-5432 obtient un score CVSS de 0.0/10, soit une gravité non évaluée. Selon le modèle EPSS, la probabilité qu'elle soit exploitée dans les 30 prochains jours est de 2.57 %, ce qui la place devant 84.9 % des CVE référencées.

Que faire ?

  • Vérifiez si SIGMA Spectrum Infusion System (versions 6.05 (model 35700BAX) with wireless battery module (WBM) version 16.) est présent sur votre parc.
  • Appliquez le correctif ou la version corrigée indiqués dans les références de l'éditeur.
  • À défaut de correctif, limitez l'exposition (filtrage réseau, désactivation du composant).
  • Automatisez la détection : les agents CVE Heimdall (gratuits, open source) corrèlent l'inventaire de vos serveurs avec la base CVE.

Références