CVE-2018-10596
Vulnérabilité élevée dans 2090 CareLink Programmer (Medtronic)
Description
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Gravité et risque d'exploitation
CVE-2018-10596 obtient un score CVSS de 7.1/10, soit une gravité élevée. Selon le modèle EPSS, la probabilité qu'elle soit exploitée dans les 30 prochains jours est de 1.32 %, ce qui la place devant 70.4 % des CVE référencées.
Que faire ?
- Vérifiez si 2090 CareLink Programmer (versions All versions) est présent sur votre parc.
- Appliquez le correctif ou la version corrigée indiqués dans les références de l'éditeur.
- À défaut de correctif, limitez l'exposition (filtrage réseau, désactivation du composant).
- Automatisez la détection : les agents CVE Heimdall (gratuits, open source) corrèlent l'inventaire de vos serveurs avec la base CVE.