CVE-2025-14095
Vulnérabilité moyenne dans ABL90 FLEX and ABL90 FLEX PLUS Analyzers (Radiometer Medical Aps)
Description
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
Gravité et risque d'exploitation
CVE-2025-14095 obtient un score CVSS de 6.8/10, soit une gravité moyenne. Selon le modèle EPSS, la probabilité qu'elle soit exploitée dans les 30 prochains jours est de 0.17 %, ce qui la place devant 5.6 % des CVE référencées.
Que faire ?
- Vérifiez si ABL90 FLEX and ABL90 FLEX PLUS Analyzers (versions All Application software versions <= 8.13 MR2, Application software versions < 6.20 MR2 with Windows 10 as underlying OS, Application software versions < 3.5MR11 with Windows 10 as underlying OS, Application software versions < 1.5.0, All application software versions with Windows 7, Windows XP as underlying OS, Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS) est présent sur votre parc.
- Appliquez le correctif ou la version corrigée indiqués dans les références de l'éditeur.
- À défaut de correctif, limitez l'exposition (filtrage réseau, désactivation du composant).
- Automatisez la détection : les agents CVE Heimdall (gratuits, open source) corrèlent l'inventaire de vos serveurs avec la base CVE.